Wanted Argon Coagulator Equipment - Buying Leads

Welcome to MedWOW’s Argon Coagulator Buying Leads section!

This section contains thousands of Argon Coagulator Buying Leads, featuring: complete systems, parts and medical supplies - all of which can be filtered by category, device, manufacturer, model, and location.

All of MedWOW’s Argon Coagulator Buying Leads are pre-qualified buyers who know exactly what they are looking for and are committed to purchase immediately. More than 10,500 visitors enter the MedWOW site daily and transactions occur quickly and on a regular basis, so Argon Coagulator leads are kept fresh and current. When you take advantage of MedWOW’s Argon Coagulator Buying Leads, you will meet new international customers who will quickly become part of your regular customers.
Applied Filters

Category : Cardiothoracic Surgery

  Reset

Device : Argon Coagulator

  Reset
No results were found for your search.
Didn't find your item? Try our other ways
Request Alert - We'll contact you when your search items are posted
Post a Buying Request - Let users from all over the world know what you are looking for
New Search - Try a different search method
 

Tips for buying Argon Coagulator

  1. These units should come with a solid-state generator, with a typical frequency range of 0.3 to 1 MHz. They should have an isolated monopolar output, a hand switch, a footswitch, and bipolar output.
  2. The output power for monopolar cutting should not exceed 300 W, and the open circuit voltage should not exceed 2,000 peak-to-peak volts; these values apply for general-purpose use.
  3. For coagulation, the output and open circuit voltage should not exceed 120 W and 6,000 peak-to-peak volts, respectively. Bipolar mode should use no more than 50 W and 300 peak-to-peak volts.
  4. Units should include the following: a return electrode contact quality monitor, independent output, visual and audible activation indicators, alarm capable of volumes higher than 45 dB at one meter, power-setting display, and convection or fan cooling system.
  5. Argon-enhanced units should have a flow range of 0 to 10 L/min, a tank capacity of more than 1,200 L (42 ft3), and a low-pressure alarm; the device should include an alarm capable of volumes higher than 45 dB at one meter.
  6. There are numerous designs of electrosurgical units, and facilities should base their comparisons among devices according to characteristics, usefulness of features, safety, and reliability - all of these affect clinical performance.
  7. Users who are considering which argon-enhanced coagulation system to purchase, should determine whether they wish to expand their current electrosurgical system with a compatible argon module or get a new one with an integral argon unit.