MedWOW / Medical Equipment / Cardiology / Cath Lab / Siemens / COROSKOP T.O.P. / Manufacturer Specifications
 

Manufacturer Specifications - COROSKOP T.O.P., Siemens

Please note: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.
MODEL

COROSKOP T,O,P,

X-ray generator (POWER NEEDED, VAC)
MODEL

POLYDOROS IS-C

GENERATORHigh frequency
kW @ 100 kVp (X-RAY GENERATOR)100
Radiographic ma (X-RAY GENERATOR)800-1,000
Radiographic kv (X-RAY GENERATOR)50-125
Radiographic timer (X-RAY GENERATOR)0.5msec-0.35 sec
Fluoroscopic ma (X-RAY GENERATOR)5-12
Fluoroscopic kv (X-RAY GENERATOR)40-125
Pulsed fluoroscopy (X-RAY GENERATOR)Yes
Cine camera range, fps (X-RAY GENERATOR)15-60
SUPPLEMENTALS (POWER REQUIREMENTS)Cardiac subtraction, angiographic subtraction, ACOM system
X-ray generator (POWER NEEDED, VAC)380.415.440. 480VAC, 3-phase
Type (CONFIGURATION)Single plane
IMAGING FEATURES
Acquisition, fps30-60
Images stored (IMAGING CHARACTERISTICS)4,000
ADDITIONAL STORAGE (NETWORKING FEATURES)ACOM T.O.P. † 
Digital imaging system (IMAGING CHARACTERISTICS)HICOR T.O.P.
TELEVISION CAMERA
GENERATORTriode-gun Saticon
Snr (TV CAMERA)56dB
Scan lines (VIDEO INPUT/OUTPUT)1,249/1,023
Bandwidth, mhz (VIDEO INPUT/OUTPUT)>25
NUMBER OF MONITORS (X-RAY TUBE)1
Monitor sizes (X-RAY TUBE)54cm
TABLETOP
Movement (TABLETOP)4-way floating
DIMENSIONS (LXW) CM, (IN) (TABLETOP)220x 43 (87 x 17)
Max. vertical movement cm (in) (TABLETOP)77.5-120(30.5-47)
Lateral (PATIENT TABLE CHARACTERISTICS)±58(±23)
Longitudinal (PATIENT TABLE CHARACTERISTICS)125(49)
Patient weight (capacity ) kg (lb) (TABLETOP)150(330)
SUPPORT (EXTERNAL BEAM)Floor
X-ray density (TABLETOP)<1 mm Al
IMAGE INTENSIFIER
Phosphor diameter Input, cm (in) (IMAGE INTENSIFIER)23/18/11(9/7/5)
Output, mm (IMAGE INTENSIFIER)23
Contrast ratio (IMAGE INTENSIFIER)20:1
Spatial resolution, center, lp/mm (IMAGE INTENSIFIER)5.2/5.8/6.2
ROOM SIZE NEEDED, M (FT) (POWER REQUIREMENTS)4.9x 6.1 (16 x 20)
X-RAY TUBE
Heat storage capacity, hu (X-RAY TUBE)1350000
Heat dissipation rate, hu/min (X-RAY TUBE)360000
Focal spot size, mm (X-RAY TUBE)0.4/0.8
Grid pulsed (X-RAY TUBE)Yes
Target angle, deg (X-RAY TUBE)8
Cooling method (X-RAY TUBE)Water
PA GANTRY
CONFIGURATIONC
Depth, cm (in) (PA GANTRY)90(35)
Rao projection (PA GANTRY)200
Lao projection (PA GANTRY)120
Cranial-to-caudal angulation, deg (PA GANTRY)+55 to -45
Sid range, cm (in) (PA GANTRY)101(40)
Motorized/manual (PA GANTRY)Motorized
Counterbalanced (PA GANTRY)Yes
SUPPORT (EXTERNAL BEAM)Floor
Park capability (PA GANTRY)Yes
OTHER SPECIFICATIONS (AIR SUPPLY)50preprogrammed projections, high- speed equipment with variable-speed movements; collimation and filter placement on LIH; motorized peripheral position for complete head-to-toe coverage.
MARKETING REGION (Interference compensation)Worldwide