MedWOW / Medical Equipment / Imaging / CT Scanner / Picker / PQ 2000 / Manufacturer Specifications
 

Manufacturer Specifications - PQ 2000, Picker

Please note: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.
MODEL

PQ 2000

X-RAY TUBE
X-RAY TUBE ANODE
Heat storage, hu (X-ray tube anode)4000000
Heat dissipation rate, hu/min (X-RAY TUBE)830000
Tube cooling (X-ray tube anode)Oil/air
Tube focal spot, mm (X-ray tube anode)0.6x 1.3; 0.9x 1.9, dual
Optional x-ray tubes (X-ray tube anode)Rhino 5,000,000 HU; Rhino 6,500,000 HU
POWER NEEDED480VAC nominal, 75kVA, 3 phase
Type (CONFIGURATION)Single slice helical
# of slices (X-ray tube anode)1
FLOOR SPACE NEEDED
Size, m2 (FLOOR SPACE NEEDED)26
Output, kw (X-RAY GENERATORS)30
Kvp range (GENERATOR)80
Ma range (IMAGING SYSTEM)30
PERFORMANCE
Minimum interscan time, sec (PERFORMANCE)1
Dynamic scan pace (PERFORMANCE)32scans/min
High-contrast spatial resolution 0%MTF, lp/cm (PERFORMANCE)21.5
50%MTF, lp/cm (PERFORMANCE)10
Low-contrast resolution, mm at %at £4 rads (PERFORMANCE)2mm at 0.35%, 1 sec scan, 6" CatPhan
Noise, %at £2.5 rads (PERFORMANCE)0.14,6.25" water phantom
TABLE FEATURES
Range of movement
Vertical, cm (Range of movement)56-103
Longitudinal, cm (Range of movement)180
Scanning range,cm (Range of movement)162
Max load capacity with accuracy, kg (Range of movement)204
HELICAL SCANNING (DISPLAY)Yes
Max scan time, sec (DISPLAY)46optional 70, 90
Max scan volume, cm (DISPLAY)162
Spatial resolution, lp/cm (DISPLAY)21.5
Pitch (DISPLAY)0.5:1 to 2:1
Reconstruction time per image, sec (DISPLAY)3
MODEL

PICKER PQ 2000

DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC)Yes
CT SIMULATION (DISPLAY)Optional
DISPLAY
MONITOR20"
Matrixes, pixels (DISPLAY)1024x 1024
Range of ct numbers (DISPLAY)-2,048to +6,143
Image enlarging scale (DISPLAY)2x to 16x
Max # slices displayed simultaneously (DISPLAY)25
3-D PACKAGE (DISPLAY)Optional
IMAGE PROCESSING (MAXIMUM DISPLAY)
Cpu (NETWORK/DATABASE)Motorola/SPARC
Scan fovs, cm (IMAGE PROCESSING)24.48
Image reconstruction matrixes (IMAGE PROCESSING)512x 512
Reconstruction time
Per slice, sec (Reconstruction time)3
For localization scan, sec (Reconstruction time)2-5
GENERATOR
Image storage
Hard disk, MB4GB; up to 29 GB optional
No. online images (IMAGE STORAGE)2,400uncompressed; up to 15,000 uncom- pressed optional
Archive (IMAGE STORAGE)8mm cartridge tape; optional optical disk
CT ANGIOGRAPHY (DISPLAY)Yes
GANTRY
Geometry (GANTRY)Rotate-rotate, slip ring
Number and type of detectors (GANTRY)4,800,solid-state cadmium tungstate
# detection channels (GANTRY)4800
Rotation times, sec 360 (GANTRY)1,1.5,2,3,4; 0.85optional
Partial (GANTRY)0.6(230°)
Slice thickness, mm (GANTRY)1,1.5,2,3,4,5, 8.1
X-ray fan beam angle, ° (GANTRY)44
Gantry angle deg (GANTRY)-30to +30
Gantry size, hxwxd,cm (GANTRY)196x 226 x 86
Gantry weight, kg (GANTRY)914.1
Gantry opening, cm (GANTRY)70
Scan localizer (GANTRY)Laser
OTHER ATTRIBUTES (Interference compensation)EL touchscreen; 3-D; volumetric MPR; bone mineral analysis; autovoice; auto- archive; autolights; autofilming; auto- window; batch filming; Toothpix dental package; digital laser inter- face; Ethernet; Voxel Q system (with CT/MR angiography).
FDA CLEARANCE (Interference compensation)Yes
CE MARK (MDD) (Interference compensation)Yes
MARKETING REGION (Interference compensation)Worldwide
You may also be interested in the following Manufacturer Specifications:
Disclaimer

Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.

MedWOW disclaims any obligation, responsibility or liability which concerns or relates to any information on this page.

CT Scanner, Picker, PQ 2000
View More Items