| MODEL | Aquilion 4 |
| Number of slices | 4 |
| SCANNER GANTRY | |
| Generation (SCANNER GANTRY) | 3rd |
| Slipring (SCANNER GANTRY) | Low voltage |
| Aperture (cm) (SCANNER GANTRY) | 72 |
| Scan fields of view (cm) (SCANNER GANTRY) | 18,24,32,40,50 |
| Nominal slice widths for axial scans (mm) (SCANNER GANTRY) | 0.5,1,2,3,4,6,8(all 4-slice mode) |
| Tilt range (degrees) (SCANNER GANTRY) | ±30 |
| Type of positioning lights (SCANNER GANTRY) | Laser |
| Accuracy of positioning lights (mm) (SCANNER GANTRY) | ±1 |
| COUCH | |
| Couch top | |
| MATERIAL | Carbon fibre |
| Length and width (cm) (SCANNER GANTRY) | 218.7 (std) or |
| Horizontal movement (Camera Operation Stand) | 188.7 (short) x 47 |
| Horizontal movement range (cm) (SCANNER GANTRY) | 219 (std) |
| Horizontal movement speeds (mm/sec) (Horizontal movement) | 186 (short) |
| Accuracy/reproducibility of table positioning (mm) (Horizontal movement) | ±0.25 |
| Scannable horizontal range (cm): | |
| (i) without table top extension (Scannable horizontal range (cm):) | 186 (std) |
| (ii) with table top extension(s) (Scannable horizontal range (cm):) | 219 (std) |
| VERTICAL MOVEMENT | 186 (short) |
| Vertical movement range out of gantry (cm) (SCANNER GANTRY) | 31-95.4 |
| Vertical movement range in gantry (cm) (Vertical movement) | 77.9-95.4 |
| Minimum couch top height outside gantry (cm) (Vertical movement) | 31 |
| Weight bearing properties | |
| Maximum weight allowed on couch (kg) (COUCH) | 205 |
| Maximum weight on couch which still achieves stated performance specifications (kg) (Weight bearing properties) | 205 |
| Type (CONFIGURATION) | High frequency |
| Location (COMPOSITION) | Rotation assembly |
| Power rating (kW) (Weight Bearing Properties) | 60 |
| kV settings available (Weight bearing properties) | 80.100.120.135 |
| mA range and step size (Weight bearing properties) | 10-500(10mA steps) |
| Max. mA allowed for each kV (Weight bearing properties) | 80 kV: 500mA |
| X-RAY TUBE | |
| Type and make (X-RAY TUBE) | Toshiba Megacool |
| Focal spot size(s) (mm), quoted to IEC 336/93 standard (X-RAY TUBE) | 0.9 x 0.8 |
| Total filtration (inherent + beam shaping filter) at central axis(mm Al equivalent) (X-RAY TUBE) | > 1 (inherent) |
| Anode heat capacity (MHU) (SCANNER GANTRY) | 1.5 - 10 (wedge dependent) |
| Maximum anode cooling rate (kHU/min) (SCANNER GANTRY) | 1386 |
| Method of cooling (X-RAY TUBE) | Oil/forced air |
| Guaranteed tube life (SCANNER GANTRY) | 300,000 rotations |
| DETECTION SYSTEM | |
| Detector type (DETECTOR CHARACTERISTICS) | Solid state |
| Number of detectors per row (Maximum anode cooling rate) | 896(plus 1 pair reference detectors at tube side) |
| Number of elements along z-axis (Maximum anode cooling rate) | 34 |
| Effective length of each element at isocentre (mm) (SCANNER GANTRY) | 4 x 0.5 30 x 1.0 |
| Total effective length of detector array at isocentre (mm) (SCANNER GANTRY) | 32 |
You may also be interested in the following Manufacturer Specifications:
- Aquilion Super4 Manufacturer Specification
- Asteion Super 4 Manufacturer Specification
- Aquilion 16 CFX Manufacturer Specification
Disclaimer
Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.
CT Scanner, Toshiba, Aquilion Multi 4
![]() | View More Items | ![]() |