MODEL

CODEMASTER 100

Hr alerts (ECG MONITOR)Configurable
Freq response, hz (ECG MONITOR)0.5-40; selectable on 12-lead 0.05-150
Lead-fault indicMessage and dotted line
BATTERY/LINE POWER (DEFIBRILLATOR)Both
Number and type (DEFIBRILLATOR)1Ni-Cd
Integral or removeRemovable
Charging system (DEFIBRILLATOR)Charger or trickle charge w/AC module
CHARGE TIME2hr to 100%
Operating time (POWER NEEDED)2.5hr monitoring or 50 max energy discharges
Dimensions (hxwxd) cm (in) (GENERATOR)8.4x 7.6 x 18.5 (3.3x 3 x 7.3)
WEIGHT, kg (lb) (DISPLAY)9.8(21.6)
Dimensions (hxwxd) cm (in) (GENERATOR)15.9x 34.9 x 38.4 (6.3x 13.7 x 15.1)
WEIGHT, kg (lb) (DISPLAY)9.8(21.6)
External outputs (UNIT)1V ECG
DEFIBRILLATOR
Energy selection, J Internal (DEFIBRILLATOR)2,3,5,7,10,20, 30.5
External (DEFIBRILLATOR)max energy > 350 (j)2,3,5,7,10,20, 30,50,70,100, 150,200,300,360
Paddle controls (DEFIBRILLATOR)Charge and discharge only
Waveform shape (DEFIBRILLATOR)Damped sinusoidal
Synchronizer (DEFIBRILLATOR)Automatic
Pediatric paddles (DEFIBRILLATOR)Integrated
Dispos electrodes (DEFIBRILLATOR)Multifunction
ECG RECORDERThermal array
Recording speed, mm/sec25
Auto/manual printBoth
Data recorded (RESERVOIR)Date, time, HR, ECG mode, event marker, defibrillator mode, selected energy, delivered energy, peak current, patient impedance, SpO2, and alarm violations (option)
Summary featureYes
EXTERNAL PACEMAKERYes
Pacing methodDemand or fixed
Pacing rate, ppmpacing rate ≤18040-180
Output current, ma36800
Pulse width, ms (STIMULUS SPECIFICATIONS)20
ECG MONITOR
Monitor type (ECG MONITOR)Integral CRT
Screen size, cm (in) (ECG MONITOR)12.7 (5)diag
Sweep speed, mm/sec (ECG MONITOR)25
Lead configuration (ECG MONITOR)I, II, III, paddles, pads, AVF, AVL, AVR, V1-V6
Through-the-paddles monitoring (ECG MONITOR)Yes
Hr display (SCAN DISPLAY)Yes
OTHER ATTRIBUTES (Interference compensation)Date, time, modes of operation, warnings and postdischarge annotation; can pace and defibrillate through same pads; charges to 360 J in <5 sec; optional high-visibility yellow; hands-free defibrillation, SpO2 with pulse oximeter; pacing; 12-lead.
FDA CLEARANCE (Interference compensation)Yes
CE MARK (MDD) (Interference compensation)Yes
MARKETING REGION (Interference compensation)Worldwide
You may also be interested in the following Manufacturer Specifications:
Disclaimer

Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.

MedWOW disclaims any obligation, responsibility or liability which concerns or relates to any information on this page.

External Defibrillator, Hewlett Packard, CodeMaster 100
View More Items