Manufacturer Specifications - Apex SPX Helix, Elscint

Please note: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.
MODEL

Apex SPX Helix

SPECT (PROCESSOR)Yes
OPTIONAL ATTACHMENTS (SHOCK-WAVE GENERATOR)Touch ruler, head support for brain SPECT, 21" non- interlaced color display
SYSTEM SPECIFICATIONS2 single-crystal detectorsdedicated camera
CollimatorsLEAP, LEUHS, LEHS, LEHR, HEAP, MEAP, LEFANHR for brain, LEAP FB for cardiac
DETECTOR/YOKE MOVEMENT
Whole-body scan
Rate, cm/min (Whole-body scan)Maximum 200 cm/min5-150
Mode (IMAGING SYSTEM)Single pass
Area, cm (Whole-body scan)54x 200
Contouring (Whole-body scan)Automatic
Yoke rotation, deg (Whole-body scan)±360
Radius, cm (in) (Whole-body scan)0-30.5(0-12)
Detector turning, deg (Whole-body scan)0-90, optional
PROCESSOR
DATA INPUTAt camera stationIntel 80486/66 MHz
Processing (PROCESSOR)Intel 80486/66 MHz
Data transfer (PROCESSOR)Floppy disk, Ethernet
Display matrix (IMAGE DISPLAY)512x 512; 1024x 1024 optional
Storage340MB hard disk; 800MB optical disk optional
HEAT LOADING, BTU/hr (gantry only) (PROCESSOR)6500
STABILITY (max °C change per hour) (PROCESSOR)3
HUMIDITY40-60%
Fwtm, ufov (Intrinsic spatial resolution, mm)7.6
Extrinsic spatial resolution (leap)
Fwhm at 10 cm, mm (Extrinsic spatial resolution (leap))8.9
Fwtm at 10 cm, mm (Extrinsic spatial resolution (leap))16.9
System sensitivity, LEAP, counts/min/μCi (Extrinsic spatial resolution (LEAP))540
Intrinsic energy resolution, fwhm, 140 kev (Extrinsic spatial resolution (leap))0.10%
Intrinsic spatial linearity, mm
Integral ufov (Intrinsic spatial linearity, mm)0.5
Differential ufov (Intrinsic spatial linearity, mm)0.1
Intrinsic uniformity, uncorrected
Integral ufov (Intrinsic spatial linearity, mm)±3.8%
Differential ufov (Intrinsic spatial linearity, mm)±2.5%
Fwhm, ufov (Intrinsic spatial resolution, mm)3.9
SYSTEM FUNCTIONS
Ufov, cm (in) (SYSTEM FUNCTIONS)54x 40 (21 x 16)
Max. count,cps (SYSTEM FUNCTIONS)500000
Dead time, μsec (SYSTEM FUNCTIONS)1.5
Energy range, kev (SYSTEM FUNCTIONS)40-520
Intrinsic spatial resolution, mm
TOTAL BODYStandard
DETECTOR CHARACTERISTICS
Crystal dimensions, cm (in) (DETECTOR CHARACTERISTICS)60x 46 (23.5x 18)
Thickness, mm (in) (DETECTOR CHARACTERISTICS)9.5(3/8)
Number of pmts (DETECTOR CHARACTERISTICS)53(3"),6(1.5")
Size, cm (in) (MAXIMUM OBJECT)7.6(3),3.8(1.5)
OPERATING PANEL
Persistence scope (OPERATING PANEL)Yes
Zoom regulation (OPERATING PANEL)Yes
RatemeterYes
Scaler (OPERATING PANEL)Yes
Time display (OPERATING PANEL)Yes
Information density (OPERATING PANEL)Yes
Energy selection
Automatic (CALIBRATION)Yes
Manual (CALIBRATION)Yes
Auto peaking (Energy selection)Yes
Image turning control (Energy selection)Yes
Pha window capacity (Energy selection)4peaks
Window selection
Automatic (CALIBRATION)Yes
Manual (CALIBRATION)Yes
Auto uniformity correction (Window selection)Yes
POWER NEEDED115VAC, 50/60 Hz, 7kVA, 3-phase
DIMENSIONS (HXWXD), CM (IN) (DETECTION PROCESS, HR)175x 100 x 130 (69x 39 x 51)
WEIGHT, kg (lb) (DISPLAY)2,586(5,700)
CLINICAL USES
OTHER ATTRIBUTES (Interference compensation)Digital camera integrated with 32-bit computer based on 80486 and an array processor; 32MB total RAM; complete clinical processing software; static, dynamic, whole-body, and multigated SPECT with circular and automatic body-contoured orbiting.
WHERE MARKETEDWorldwide
You may also be interested in the following Manufacturer Specifications:
Disclaimer

Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.

MedWOW disclaims any obligation, responsibility or liability which concerns or relates to any information on this page.

Gamma Camera, Scanning, Elscint, Apex SPX Helix
View More Items