| MODEL | Cardio 60 |
| SPECT (PROCESSOR) | Yes |
| POWER NEEDED | 208-240 VAC, 30 A |
| DIMENSIONS (HXWXD), CM (IN) (DETECTION PROCESS, HR) | 186x 168 x 439 (73x 66 x 173) |
| SYSTEM SPECIFICATIONS | 2 single-crystal detectorsdedicated camera |
| SYSTEM FUNCTIONS | |
| Ufov, cm (in) (SYSTEM FUNCTIONS) | 50.8x 38.1 (20x 15) |
| Max. count,cps (SYSTEM FUNCTIONS) | 250000 |
| Dead time, μsec (SYSTEM FUNCTIONS) | 1.3 |
| Energy range, kev (SYSTEM FUNCTIONS) | 56-920 |
| Intrinsic spatial resolution, mm | |
| HUMIDITY | 45-75% |
| Fwtm, ufov (Intrinsic spatial resolution, mm) | 6.65 |
| Extrinsic spatial resolution (leap) | |
| Fwhm at 10 cm, mm (Extrinsic spatial resolution (leap)) | 8.8 |
| Fwtm at 10 cm, mm (Extrinsic spatial resolution (leap)) | 16 |
| System sensitivity, LEAP, counts/min/μCi (Extrinsic spatial resolution (LEAP)) | 265 |
| Intrinsic energy resolution, fwhm, 140 kev (Extrinsic spatial resolution (leap)) | 0.10% |
| Intrinsic spatial linearity, mm | |
| Integral ufov (Intrinsic spatial linearity, mm) | 0.5 |
| Differential ufov (Intrinsic spatial linearity, mm) | 0.2 |
| Intrinsic uniformity, uncorrected | |
| Integral ufov (Intrinsic spatial linearity, mm) | ±2.5% |
| Differential ufov (Intrinsic spatial linearity, mm) | ±2% |
| DETECTOR/YOKE MOVEMENT | |
| Whole-body scan | |
| Rate, cm/min (Whole-body scan) | Maximum 200 cm/min5-190 |
| Mode (IMAGING SYSTEM) | Dual pass |
| Area, cm (Whole-body scan) | 51x 206 |
| Contouring (Whole-body scan) | Manual, automated learn mode |
| Yoke rotation, deg (Whole-body scan) | 370(±185) |
| Radius, cm (in) (Whole-body scan) | 11.2-37.2(4.4-14.6) |
| Detector turning, deg (Whole-body scan) | 0,fixed at 90 |
| Clearance | |
| Maximum, cm (in) (Clearance) | 74.7(29.4) |
| Minimum, cm (in) (Clearance) | 22.4(8.8) |
| TOTAL BODY | Yes |
| Fwhm, ufov (Intrinsic spatial resolution, mm) | 3.4 |
| PROCESSOR | |
| DATA INPUT | At camera stationSun SPARC based |
| Processing (PROCESSOR) | Sun Blade 1000 |
| Data transfer (PROCESSOR) | DICOM, Ethernet, In- terfile 3.3, floppy |
| Dicom 3.0 (PROCESSOR) | Yes |
| Display matrix (IMAGE DISPLAY) | 1152x 900 |
| Storage | 20GB hard disk, 1.3GB optical disk, 1.5MB tape, 1.44 MB floppy, CD drive |
| HEAT LOADING, BTU/hr (gantry only) (PROCESSOR) | 6500 |
| STABILITY (max °C change per hour) (PROCESSOR) | 2.8 |
| DETECTOR CHARACTERISTICS | |
| Crystal dimensions, cm (in) (DETECTOR CHARACTERISTICS) | 52x 64 (20.5x 25) |
| Thickness, mm (in) (DETECTOR CHARACTERISTICS) | 9.5(3/8) |
| Number of pmts (DETECTOR CHARACTERISTICS) | 49(3").6(2") |
| Size, cm (in) (MAXIMUM OBJECT) | 7.6(3),5(2) |
| Digital detectors (DETECTOR CHARACTERISTICS) | 1ADC/PMT |
| Collimators | LEGP, LEHR, MEGP, HEGP, HEHR, VXHR, VXGP, LEHS, HEGH, fanbeam |
| OPERATING PANEL | |
| Persistence scope (OPERATING PANEL) | Yes |
| Zoom regulation (OPERATING PANEL) | Yes |
| Ratemeter | Yes |
| Scaler (OPERATING PANEL) | Yes |
| Time display (OPERATING PANEL) | Yes |
| Information density (OPERATING PANEL) | Whole body |
| Energy selection | |
| Automatic (CALIBRATION) | Yes |
| Manual (CALIBRATION) | Yes |
| Auto peaking (Energy selection) | Yes |
| Image turning control (Energy selection) | Yes |
| Pha window capacity (Energy selection) | 3peaks |
| Window selection | |
| Automatic (CALIBRATION) | Yes |
| Manual (CALIBRATION) | Yes |
| Auto uniformity correction (Window selection) | Yes |
| Nonuniform attenuation correction method (Window selection) | Optional line source |
| CLINICAL USES | |
| WEIGHT, kg (lb) (DISPLAY) | 1,964(4,330) |
| OPTIONAL ATTACHMENTS (SHOCK-WAVE GENERATOR) | Vantage Attenuation Correction |
| OTHER ATTRIBUTES (Interference compensation) | EPIC digital detectors with remote diagnostics; 90° detector configuration; semiautomatic exchanger; software- upgradable detector; robotic-controlled motions. ETL listed; GMP compliant. |
| FDA CLEARANCE (Interference compensation) | Yes |
| CE MARK (MDD) (Interference compensation) | Yes |
| WHERE MARKETED | Asia, Europe, |
You may also be interested in the following Manufacturer Specifications:
- CardioMD Manufacturer Specification
- Forte with Jetstream Manufacturer Specification
- Meridian Manufacturer Specification
Disclaimer
Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.
Gamma Camera, Scanning, Philips, Vertex Cardio 60
![]() | View More Items | ![]() |