| MODEL | Symbia E |
| SPECT | Yes |
| OPTIONAL ATTACHMENTS | Pediatric and scintimammography pallets, clinical engines, cardio•Flash and onco•Flash half-time imaging and iterative reconstruction, cardiac quantification and reporting software, syngo MI mobile, and 5/8-inch crystal |
| SYSTEM SPECIFICATIONS | Dual detector, variable angle |
| POWER NEEDED | 200-240 |
| WEIGHT, kg | 1,755 (3,900), with 2 HE collimators |
| DIMENSIONS | 193 x 167.6 x 169.5 (76 x 66 x 66.8) |
| PROCESSOR | |
| Data transfer | DICOM, TCP/IP |
| HEAT LOADING, BTU/hr | 7400 |
| Dicom 3.0 | Standard |
| STABILITY | 4.4 |
| Processing | syngo MI Workplace P |
| Storage | 40 GB OS data, 80 GB patient data |
| DATA INPUT | syngo Acquisition Workplace |
| Display matrix | 1280 x 1024 |
| DETECTOR CHARACTERISTICS | |
| Thickness, mm | 9.5 (3/8) |
| Digital detectors | ADC on each PMT |
| Size, cm | 7.6 (3), 5 (2) |
| Number of pmts | 53 (3 in), 6 (2 in); per detector |
| Crystal dimensions, cm | 59.1 x 44.5 (23.3 x 17.4) |
| OPERATING PANEL | |
| Nonuniform attenuation correction | Optional profile, transmission source-based c.clear attenuation correction |
| Persistence scope | Yes |
| Ratemeter | Yes |
| Time display | Yes |
| Scaler | Yes |
| Zoom regulation | Yes |
| Pha window capacity | 6 energy windows |
| Autopeaking | Yes |
| Manual | Yes |
| Automatic | Yes |
| Auto uniformity correction | Yes |
| Automatic | Yes |
| Manual | Yes |
| Image turning control | Yes |
| SYSTEM FUNCTIONS | |
| Fwtm at 10 cm, mm | 13.6 with LEHR |
| Intrinsic energy resolution, fwhm, 140 kev | ≤9.9% |
| Integral ufov | ≤0.7 absolute |
| Energy range, kev | 50-511 |
| Max. count,cps | 310,000 without scatter |
| Fwhm, ufov | ≤3.9 |
| Differential ufov | ≤0.2 |
| Differential ufov | ≤2.7% |
| Fwhm at 10 cm, mm | 7.5 with LEHR |
| Ufov, cm | 53.3 x 38.7 (21 x 15.25) |
| Fwtm, ufov | ≤7.7 |
| Integral ufov | ≤3.7% |
| System sensitivity, leap, cpm/µci | 202 with LEHR |
| Collimators | LEHS, LEAP, LEHR, LEUHR, LEFB, ME, HE, UHE, pinhole |
| HUMIDITY | 15-80% |
| TOTAL BODY | Yes |
| DETECTOR/YOKE MOVEMENT | |
| Yoke rotation, deg | 440 CW, 30 CCW |
| Mode | Single pass |
| Contouring | Automatic |
| Radius, cm | 14-67, LEHR collimator |
| Rate, cm/min | 0.1-240 |
| Area, cm | 53.3 x 200 |
| Detector turning, deg | 20 cephalic in, 90 caudal out |
| FDG-SPECT | Yes |
| CLINICAL USES | General nuclear medicine; whole body, cardiac, and general SPECT |
| OTHER ATTRIBUTES | SPECT with 76 |
| FDA CLEARANCE | Yes |
| CE MARK | Yes |
| WHERE MARKETED | Worldwide |
You may also be interested in the following Manufacturer Specifications:
- e.cam Multiangle Cardiac Manufacturer Specification
- e.cam Variable Angle Manufacturer Specification
- e.cam Duet Manufacturer Specification
Disclaimer
Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.
Gamma Camera, Scanning, Siemens, Symbia E Single
![]() | View More Items | ![]() |