MedWOW / Medical Equipment / Nuclear Medicine / Gamma Camera, Scanning / Siemens / Symbia S / Manufacturer Specifications
Manufacturer Specifications - Symbia S, Siemens
Please note: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.

| MODEL | Symbia S |
| SPECT | Yes |
| OPTIONAL ATTACHMENTS | 5/8 in crystal option, Flash 3-D advanced SPECT interactive reconstruction, CT attenuation correction, IDL user programming, e.soft express remote viewing and processing software, dual-monitor display |
| SYSTEM SPECIFICATIONS | Dual detector, variable angle |
| Collimators | LEHS, LEAP, LEHR, MELP, HE, UHE, pinhole |
| SYSTEM FUNCTIONS | |
| Differential ufov | ≤2.7% |
| Integral ufov | ≤3.7% |
| Max. count,cps | 310,000 without scatter per detector |
| Fwhm, ufov | ≤3.8 |
| Fwtm, ufov | ≤7.7 |
| Fwtm at 10 cm, mm | 14.1 with LEHR |
| System sensitivity, LEAP, counts/min/μCi | 202 with LEHR per detector; 565 (5/8 in) with MELP |
| Ufov, cm | 53.3 x 38.7 (21 x 15.3) |
| Intrinsic energy resolution, fwhm, 140 kev | ≤9.9% |
| Integral ufov | ≤0.7 absolute |
| Fwhm at 10 cm, mm | 7.4 with LEHR |
| Differential ufov | ≤0.2 |
| Energy range, kev | 50-511 |
| POWER NEEDED | 200/208/220/240 VAC, 47-63 Hz, single phase |
| WEIGHT, kg | 2,624 (5,784), with 2 HE collimators |
| DIMENSIONS | 225 x 193 x 225 (88.7 x 76 x 88.6) |
| OPERATING PANEL | |
| Persistence scope | Yes |
| Time display | Yes |
| Nonuniform attenuation correction | Optional, CT-based, field upgradable to TruePoint SPECT/CT |
| Image turning control | Yes |
| Pha window capacity | 6 energy windows |
| Autopeaking | Yes |
| Manual | Yes |
| Ratemeter | Yes |
| Automatic | Yes |
| Scaler | Yes |
| Zoom regulation | Yes |
| Auto uniformity correction | Yes |
| Manual | Yes |
| Automatic | Yes |
| HUMIDITY | 15-80% |
| DETECTOR/YOKE MOVEMENT | |
| Mode | Single pass |
| Yoke rotation, deg | 405 CW, 135 CCW |
| Minimum, cm | 11.9 (4.7) HE |
| Maximum, cm | 73.2 (28.8) LEHR |
| Rate, cm/min | 3-600 |
| Contouring | Automatic |
| Area, cm | 53.3 x 200 |
| Detector turning, deg | +/-16 |
| PROCESSOR | |
| Display matrix | 1280 x 1024 |
| DATA INPUT | e.soft A |
| Processing | e.soft P |
| Dicom 3.0 | Standard |
| STABILITY | 4.4 |
| Storage | 40 GB OS data, 80 GB patient data |
| HEAT LOADING, BTU/hr | 6,800 |
| Data transfer | DICOM, TCP/IP |
| DETECTOR CHARACTERISTICS | |
| Number of pmts | 53 (3 in), 6 (2 in); per detector |
| Crystal dimensions, cm | 59.1 x 44.5 (23.3 x 17.4) |
| Size, cm | 7.6 (3), 5 (2) |
| Thickness, mm | 9.5 (3/8), 15.9 (5/8) |
| Digital detectors | ADC on each PMT |
| FDG-SPECT | Yes |
| TOTAL BODY | Yes |
| PLANNING & PURCHASE | |
| CLINICAL USES | General nuclear medicine, whole body cardiac and general SPECT; simultaneous dynamic and planar static acquisition, simultaneous gated and ungated SPECT acquisition, dynamic SPECT mode |
| OTHER ATTRIBUTES | SPECT with 76 |
| FDA CLEARANCE | Yes |
| CE MARK | Yes |
| WHERE MARKETED | Worldwide |
You may also be interested in the following Manufacturer Specifications:
- e.cam Multiangle Cardiac Manufacturer Specification
- e.cam Variable Angle Manufacturer Specification
- e.cam Duet Manufacturer Specification
Disclaimer
Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.
Gamma Camera, Scanning, Siemens, Symbia S
![]() | View More Items | ![]() |