MODEL

CardioMD

SPECT (PROCESSOR)Yes
OPTIONAL ATTACHMENTS (SHOCK-WAVE GENERATOR)AUTOQUANT, CardiaQ, Emory Toolbox
SYSTEM SPECIFICATIONS2 single-crystal detectorsdedicated camera
CollimatorsLEGP, LEHR
HUMIDITY45-80%
PROCESSOR
DATA INPUTMobile acquisition consoleWin-based
Processing (PROCESSOR)Jetstream, Windowsbased
Data transfer (PROCESSOR)DICOM, Ethernet, TCP/IP, CD-RW
Dicom 3.0 (PROCESSOR)Yes
Display matrix (IMAGE DISPLAY)1280 x 1024
Storage80 GB hard disk, CD-RW, 1.44 MB floppy (minimum)
HEAT LOADING, BTU/hr (gantry only) (PROCESSOR)<4,800
STABILITY (max °C change/hour) (PROCESSOR)5
DETECTOR CHARACTERISTICS
Crystal dimensions, cm (in) (DETECTOR CHARACTERISTICS)40.2 x 24.6 (15.8 x 9.7)
Thickness, mm (in) (DETECTOR CHARACTERISTICS)9.5 (3/8)
Number of pmts (DETECTOR CHARACTERISTICS)24
Size, cm (in) (MAXIMUM OBJECT)7.6 (3), square PMT
Digital detectors (DETECTOR CHARACTERISTICS)Yes
SYSTEM FUNCTIONS
Ufov, cm (in) (SYSTEM FUNCTIONS)36 x 20.4 (14.2 x 8)
Max. count,cps (SYSTEM FUNCTIONS)200000
Dead time, μsec (SYSTEM FUNCTIONS)0.7
Energy range, kev (SYSTEM FUNCTIONS)60-170
Intrinsic spatial resolution, mm
Fwhm, ufov (Intrinsic spatial resolution, mm)3.7
Fwtm, ufov (Intrinsic spatial resolution, mm)7.6
Extrinsic spatial resolution (leap)
Fwhm at 10 cm, mm (Extrinsic spatial resolution (leap))9.4
System sensitivity, leap, counts/min/uci (Extrinsic spatial resolution (leap))277
Intrinsic energy resolution, fwhm, 140 kev (Extrinsic spatial resolution (leap))0.10%
Intrinsic spatial linearity, mm
Integral ufov (Intrinsic spatial linearity, mm)0.5
Differential ufov (Intrinsic spatial linearity, mm)0.2
Intrinsic uniformity, uncorrected
Integral ufov (Intrinsic spatial linearity, mm)±2.5%
Differential ufov (Intrinsic spatial linearity, mm)±1.5%
POWER NEEDED100/120 VAC ±10%, 12 A; 200/220/240 VAC ±10%, 6.25 A
DIMENSIONS (HXWXD), CM (IN) (DETECTION PROCESS, HR)186 x 189 x 200 (73 x 74 x 79)
WEIGHT, kg (lb) (DISPLAY)1,100 (2,426)
DETECTOR/YOKE MOVEMENT
Whole-body scan
Contouring (Whole-body scan)Automatic
Yoke rotation, deg (Whole-body scan)180 (CCW, CW)
Radius, cm (in) (Whole-body scan)32.3-52.8 (12.7-20.8)
Detector turning, deg (Whole-body scan)-45 to +45 supine, -45 to +135 prone
OPERATING PANEL
Persistence scope (OPERATING PANEL)Yes
Zoom regulation (OPERATING PANEL)Yes
RatemeterYes
Scaler (OPERATING PANEL)Yes
Time display (OPERATING PANEL)Yes
Energy selection
Automatic (CALIBRATION)Yes
Manual (CALIBRATION)Yes
Auto peaking (Energy selection)Yes
Image turning control (Energy selection)Yes
Pha window capacity (Energy selection)3 peaks
Window selection
Automatic (CALIBRATION)Yes
Manual (CALIBRATION)Yes
Auto uniformity correction (Window selection)Not required
Nonuniform attenuation correction (Window selection)Optional
CLINICAL USESCardiology
OTHER ATTRIBUTES (Interference compensation)Integrated gantry/ table; self-tuning detector.
FDA CLEARANCE (Interference compensation)Yes
CE MARK (MDD) (Interference compensation)Yes
WHERE MARKETEDAsia, Europe, Latin America, North America