Manufacturer Specifications - IWS 3500/Panda 3500 Bassinet, Ohmeda (GE Healthcare)

Please note: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.
MODEL

IWS 3500_Panda 3500 Bassinet

TEMPERATURE
Range, °c (TEMPERATURE)Below or equal 45 °C30-42
Set-point display (TEMPERATURE)LED display
Skin-temp display (TEMPERATURE)LED
WHEELS4
Size, cm (in) (MAXIMUM OBJECT)12.7 (5)
Apgar timer (WHEELS)Yes
Drawers/shelves (WHEELS)Optional
BRAKES2
Phototherapy lights (WHEELS)Optional
Resuscitation module (WHEELS)Optional
Type/number (ANALYZER)Halogen/1
CONFIGURATIONFreestanding
ALERTS (CONFIGURATION)Yes
ToneProgressive 1
Alert signalAudible and visual
Manual modeSingle tone after 12 min at <25% power, 2-tone alarm at 3 min, heater off, See Other Specs
Sensor disconnectAudible and visual
Power loss (ALERTS)Audible and visual 2
Trigger, ±°c from set pointFactory set at 1, adjustable to 0.5 Progressive alarm begins with a single-tone user prompt; if it is not silenced in 30 sec, it progresses to a more urgent tone. After 3 min, it progresses to a 2-tone alarm.
BASSINET (CONTROL CONSOLE)Yes
L x w, cm (in) (CONTROL CONSOLE)69 x 51 (27 x 20)
Height from floor, cm (in) (CONTROL CONSOLE)117.5 (46.25)
Mattress surface tilt, ° (CONTROL CONSOLE)8 trend, 4 rev trend
Distance from heat source, cm (in) (CONTROL CONSOLE)69 (27)
Sidewall, cm (in) (CONTROL CONSOLE)Yes
HEAT SOURCE (CONTROL CONSOLE)540 W cal rod
SELF-CHECK FEATURES (CONTROL CONSOLE)Power-up checks: calibration, control system, software, solid-state relay gives simulated alarm; continuous checks: software, calibration 3
Illumination (DETECTOR)Bulbs
POWER NEEDED, VAC (ENERGY)100/120/220/230/240
DIMENSIONS (HXWXD) CM, (IN) (DISPLAY)185 x 62 x 95 (72.8 x 24.4 x 37.4)
WEIGHT, kg (lb) (DISPLAY)98 (215)
CONTROL CONSOLE
Heat signal (CONTROL CONSOLE)Bar graph
HEAT CONTROL (CONTROLS)Automaticand manual
Range, °c (TEMPERATURE)Below or equal 45 °C35-37.5
OTHER ATTRIBUTES (Interference compensation)Panda Turbo pre-heat to 100% power for 12 min then to 50% power with no alarms until mode activation then alarm structure same as manual mode (see Alarms above). Meets requirements of IEC 601-2-21 and ISO 9000.
FDA CLEARANCE (Interference compensation)Yes
CE MARK (MDD) (Interference compensation)Yes
MARKETING REGION (Interference compensation)Worldwide
You may also be interested in the following Manufacturer Specifications:
Disclaimer

Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.

MedWOW disclaims any obligation, responsibility or liability which concerns or relates to any information on this page.

Infant Warmer, Ohmeda (GE Healthcare), IWS 3500/Panda 3500 Bassinet
View More Items