MODEL

Flo-Gard 6301

TOTAL VOLUME INFUSED
Display, ml (TOTAL VOLUME INFUSED)1-9,999
KVO RATE, mL/hr (TOTAL VOLUME INFUSED)5-6
POWER NEEDED, VAC (ENERGY)115
Duration of recharge,hr (TOTAL VOLUME INFUSED)80% @ 8
Frequency, hz (REVOLUTIONS)60
Battery life (TOTAL VOLUME INFUSED)6 @ 1,400; 4 @ 1,400 using both channels
DIMENSIONS (HXWXD) (RISE TIME, msec)29 x 32.5 x 13 (11.4 x 12.8 x 5.1)
WEIGHT, kg (lb) (DISPLAY)8.1
FLUID RESISTANT (TOTAL VOLUME INFUSED)Yes
AUDIBLE ALARM
Alarm pause (CONFIGURATION)Yes
VOLUME REGULATION (POWER NEEDED)Yes
AUTOMATIC PIGGYBACKING (ALERTS AND SIGNALS)Yes
PUMP SYSTEM (AUDIBLE ALARM)Peristaltic type - for details please see comments.Linear peristaltic
ALERTS AND SIGNALS
Air in line (ALERTS AND SIGNALS)Yes
Set disengaged (ALERTS AND SIGNALS)Yes
Occlusion downstream (ALERTS AND SIGNALS)Yes
Occlusion upstream (ALERTS AND SIGNALS)Yes
Circuit failure (ALERTS AND SIGNALS)Yes
Infusion terminated (ALERTS AND SIGNALS)Yes
User prompts (ALERTS AND SIGNALS)Yes
Empty battery (ALERTS AND SIGNALS)Yes
Detection mechanism (ALERTS AND SIGNALS)Benchmarking/selectable
Door open (ALERTS AND SIGNALS)Yes
Depleted fluid container (ALERTS AND SIGNALS)No, VTBI
Pressure, psi (ALERTS AND SIGNALS)Selectable, 3 levels of sensitivity
Low battery (EQUIPMENT ALARMS)Yes
DATA LOG (ALERTS AND SIGNALS)Yes
IV SET
Non-needle iv connection (IV SET)Yes
Kind of set (IV SET)Baxter
Free-flow safeguard (IV SET)Optional - for details please see comments.Optional, depends on configuration
NURSE CALL JACK (ALERTS AND SIGNALS)Yes
Number of channels (RECORDER)2
FLOW RANGE, mL/hr (AUDIBLE ALARM)1-99.9, 1-1,999, in 0.1 mL/hr and 1 mL/hr increments, respectively
Accuracy, % (POWER NEEDED)5
VTBI SELECTOR, mL (AUDIBLE ALARM)1-9,999
OTHER ATTRIBUTES (Interference compensation)Remote monitoring; postocclusion restart; front-panel indication of flow resistance; selectable options for customized operation; uses standard Baxter IV sets.
Supplier Footnotes (PURCHASE INFORMATION)<1>Specifications current as of March 2004.
FDA CLEARANCE (Interference compensation)Yes
MARKETING REGION (Interference compensation)Worldwide