Tips for buying Cardiac Ablation System

  1. Ablation technologies are combined methods of multiple modalities that are individually used to excise, destroy or otherwise effectively treat tissue therapeutically for the various purposes of removing dead tissue or other biological matter.
  2. To ensure safe and efficient operation of cardiac ablation systems and interstitial thermotherapy systems, facilities need to look into several considerations including: RF generator output, which should be between 0 to 50 W; the waveform should be approximately 500 sinusoidal; stimulator output should range from 0 to 10 V and 0 to 10 mA.
  3. Pulse duration of a cardiac ablation system should be between 0 to 3 msec, with a frequency of 2 to 200 Hz.
  4. To ensure suitable functioning during its operation, the cardiac ablation system should display the current and voltage range, impedance range, and temperature range.
  5. It is recommended to have both visual and audio indicators to show activation. The cardiac ablation system should have a self-test mode and include protective circuits such as patient isolation, fault conditions, and impedance and temperature monitors.