MedWOW / Medical Equipment / Mobile Units / Mobile MRI / Hitachi / Mobile MRP-7000 / Manufacturer Specifications
 

Manufacturer Specifications - Mobile MRP-7000, Hitachi

Please note: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.
MODEL

MRP7000

MAGNETPermanent
POWER NEEDED
LINE VOLTAGE, VAC190-240, single- phase
Kva (POWER NEEDED)7.5
CLINICAL USE (POWER NEEDED)Whole body
SURFACE COILS
Spine (SURFACE COILS)Med & lg QD flex
Knee (SURFACE COILS)Latchable QD
Neck (SURFACE COILS)Latchable solenoid
Tmj (SYSTEM CAPABILITIES)Opt bilat solenoid
Extremity (SURFACE COILS)Ltchbl oval solenoid
Head (SURFACE COILS)QD
Breast (APPLICATORS)Opt bilateral QD
Shoulder (SURFACE COILS)Latchable solenoid
Additional (Alarms, high/low)Optional QD pelvis, QD C-spine, small & extra-large QD flex, small extremity
BORE SIZE, dia x L
Or w x h x l, cm (BORE SIZE, dia x L)101x 46, wide bore
DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC)Storage class
AngiographyYes
CARDIAC GATING
Ecg/peripheral (CARDIAC GATING)Yes/yes
IMAGING
Pulse sequences (IMAGING)SE, GE, GR, IR, FIR, STIR, FSE, FLAIR, dual slice, SARGE, RF spoiled/rephase SARGE, 2-D/3-D TOF, CE-MRA, SSP, MTC, MRCP, MR fluoro, segmented K-space
Repetition time, msec (IMAGING)15-7,000
Echo time, msec (IMAGING)4-402
Inversion time, msec (IMAGING)80-2,000
Slice thickness, mm (GANTRY)3-50 (2-D), 0.5-5 (3-D)
Fov, cm (IMAGING)12-42
Fov offsets (IMAGING)Up to 21 cm
Scan orientations (IMAGING)Transverse, coronal, sagittal, single oblique, multiple oblique angles
Measuring matrix (IMAGING)(128-512, steps of 15,phase) x (256- 512freq); 1/2 scan, rectangular
Display matrix (IMAGE DISPLAY)512x 512
Pixel intensity (IMAGING)-2,000to +4,000
Resolution, mm (PERFORMANCE)0.4
OPTIONSFS package, MRA, UPS, VCR interface and joint motion packages, remote monitor, laser imager
BORE FEATURES (DIAMETER, cm)Wide bore, short depth, intercom, SoftSound gradients
MAGNET WEIGHT, kg (CRYOGEN USE, L/hr)16000
DIMENSIONS (HXWXD),CM (CRYOGEN USE, L/hr)175x 228 x 160
IMAGING MODES
Single (EYEPIECE)Yes
Multislice (IMAGING MODES)Yes
Volume study (IMAGING MODES)Yes
Additional (Alarms, high/low)Multislab, multiangle, multiple slice/multiple angle, FSE, dual slice, MRA, gating
Reconstruction time
Single slice, sec (RECONSTRUCTION TIME)<2
Multislice, sec (RECONSTRUCTION TIME)<2/slice
Volume, sec (RECONSTRUCTION TIME)<2/slice
ELECTROMAGNETIC FIELD STRENGTH, T (5-GAUSS FRINGE FIELD)0.3
GRADIENT SUBSYSTEM
Strength, mt/m (GRADIENT SUBSYSTEM)8
Rise time to 10mT/m, msec (GRADIENT SUBSYSTEM)<1
Cpu (NETWORK/DATABASE)Motorola 32-bit
Memory size, mb (GRADIENT SUBSYSTEM)16
ARRAY PROCESSOR (EXTERNAL BEAM)Hitachi
Memory size, mb (GRADIENT SUBSYSTEM)126
Storage media/size (GRADIENT SUBSYSTEM)Magnetic disk, optical disk
Image storage capacity (GRADIENT SUBSYSTEM)5,000magnetic disk, 19,000optical disk
SHIMMING (CRYOGEN USE, L/hr)Passive & 3-axis per patient auto shim
Diameter, cm (VACUUM)
Body coil (DIAMETER, cm)33.38.48.61; flexible quadrature
Head coil (DIAMETER, cm)27x 22 quadrature
5-GAUSS FRINGE FIELD
Radial/axial, m (5-GAUSS FRINGE FIELD)<2.5/2.1
CONFIGURATIONOpen
FDA CLEARANCE (Interference compensation)Yes
CE MARK (MDD) (Interference compensation)No
MARKETING REGION (Interference compensation)Asia, Europe, North
You may also be interested in the following Manufacturer Specifications:
Disclaimer

Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.

MedWOW disclaims any obligation, responsibility or liability which concerns or relates to any information on this page.

Mobile MRI