MODEL

E scan XQ

MAGNETPermanent
POWER NEEDED
LINE VOLTAGE, VAC110/220
Kva (POWER NEEDED)0.3
A/c, btu/hr (POWER NEEDED)4500
CLINICAL USE (POWER NEEDED)Extrem/shoulder/hip
GRADIENT SUBSYSTEM
Strength, mt/m (GRADIENT SUBSYSTEM)20
Rise time to 10 mt/m, msec (GRADIENT SUBSYSTEM)0.8
Cpu (NETWORK/DATABASE)Pentium III
Memory size, mb (GRADIENT SUBSYSTEM)320 + 256 KB cache
ARRAY PROCESSOR (EXTERNAL BEAM)Digital signal processor
Memory size, mb (GRADIENT SUBSYSTEM)128.5
Storage media/size (GRADIENT SUBSYSTEM)650 MB optical disk, 650 MB CD-R
Image storage capacity (GRADIENT SUBSYSTEM)4500
BORE FEATURES (DIAMETER, cm)Self-centering and rotating coils, table rotates to 6 positions for optimal patient access
OPTIONSRF shielding pavilion
ELECTROMAGNETIC FIELD STRENGTH, T (5-GAUSS FRINGE FIELD)0.2
DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC)Yes
5-GAUSS FRINGE FIELD
Radial/axial, m (5-GAUSS FRINGE FIELD)1.3/1.5
SURFACE COILS
Knee (SURFACE COILS)Dual phased array
Extremity (SURFACE COILS)Flexible coil (hip)
Shoulder (SURFACE COILS)Solenoid, 2 sizes
Additional (Alarms, high/low)Dual phased array ankle and wrist coil
BORE SIZE, dia x L
Or w x h x l, cm (BORE SIZE, dia x L)24-30, open
IMAGING
Pulse sequences (IMAGING)Multiplanar scout, SE, GE, ME, STIR, TSE, TME, HSE, HFE, T 3-D, GEFS, GE 3-D, TGE
Repetition time, msec (IMAGING)5-00
Echo time, msec (IMAGING)8.2
Inversion time, msec (IMAGING)20-800
Slice thickness, mm (GANTRY)0.6-10
Fov, cm (IMAGING)10-30, 14 displayed
Scan orientations (IMAGING)Axial, coronal, sagittal, oblique, double oblique
Measuring matrix (IMAGING)Selectable to 512 x 512
Display matrix (IMAGE DISPLAY)510 x 478
Pixel intensity (IMAGING)0-4,096
Resolution, mm (PERFORMANCE)0.4
IMAGING MODES
Single (EYEPIECE)Yes
Multislice (IMAGING MODES)Yes
Volume study (IMAGING MODES)Yes
Additional (Alarms, high/low)3-plane scout view (localizer)
Reconstruction time
Single slice, sec (RECONSTRUCTION TIME)0.8
Multislice, sec (RECONSTRUCTION TIME)0.8/slice
Volume, sec (RECONSTRUCTION TIME)0.8/slice
SHIMMING (CRYOGEN USE, L/hr)Passive
Cooling method (X-RAY TUBE)Not required
MAGNET WEIGHT, kg (CRYOGEN USE, L/hr)1930
DIMENSIONS (HXWXD),CM (CRYOGEN USE, L/hr)150 x 125 x 150
CONFIGURATIONOpen
OTHER ATTRIBUTES (Interference compensation)Dedicated open unit for imaging extremities, including shoulders; self-centering RF coil; patient support allows free positioning; fits in 200 ft2 space.
FDA CLEARANCE (Interference compensation)Yes
CE MARK (MDD) (Interference compensation)Yes
MARKETING REGION (Interference compensation)Worldwide (Esaote), USA (GE Healthcare)