MODEL

MAGNETOM Vision

ELECTROMAGNETIC FIELD STRENGTH, T (5-GAUSS FRINGE FIELD)1.5
YEAR FIRST INSTALLED (Storage)1994
CLINICAL USE (POWER NEEDED)Whole body
Respiratory gating (DISPLAY)Yes
CARDIAC GATING
Ecg/peripheral (CARDIAC GATING)Yes/yes
GRADIENT SUBSYSTEM
Strength, mt/m (GRADIENT SUBSYSTEM)25
Rise time to 10mT/m, msec (GRADIENT SUBSYSTEM)<0.24; <0.12 opt
Cpu (NETWORK/DATABASE)Sun-based host integrated into SMI
Memory size, mb (GRADIENT SUBSYSTEM)64
ARRAY PROCESSOR (EXTERNAL BEAM)Siemens Medical Imager (SMI)
Memory size, mb (GRADIENT SUBSYSTEM)576
Storage media/size (GRADIENT SUBSYSTEM)5.25" MOD drive; 600MB
Image storage capacity (GRADIENT SUBSYSTEM)>9,000 (256 x 256)
SPECTROSCOPY (GRADIENT SUBSYSTEM)Optional1H Package
DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC)Optional
OPTIONSPACSnet Ethernet connection, color monitor, satellite console, Prominence workstation, SatPads for uniform coil load, Advanced coil- positioning kit, additional memory & optical disk drive
SURFACE COILS
Spine (SURFACE COILS)CP array std
Knee (SURFACE COILS)CP trans/rec std
Neck (SURFACE COILS)HH volume std
Tmj (SYSTEM CAPABILITIES)Bilateral option
Extremity (SURFACE COILS)CP Flex pkg, option
Head (SURFACE COILS)CP std
Breast (APPLICATORS)Bilateral option
Shoulder (SURFACE COILS)CP Flex pkg option
Additional (Alarms, high/low)Head/neck MRA, endo- rectal, CP body array and movable spine
BORE DIAMETER or W x H, cm (SURFACE COILS)Round, normal size.61W, 54 diagonal
IMAGING MODES
Single (EYEPIECE)Yes
Multislice (IMAGING MODES)Yes
Volume study (IMAGING MODES)Yes
5-GAUSS FRINGE FIELD
Radial/axial, m (5-GAUSS FRINGE FIELD)2.6/4.7
Reconstruction time
Single slice, sec (RECONSTRUCTION TIME)<0.1
Multislice, sec (RECONSTRUCTION TIME)<0.1
Volume, sec (RECONSTRUCTION TIME)<0.1
SHIMMING (CRYOGEN USE, L/hr)Passive, active
Diameter, cm (VACUUM)
Body coil (DIAMETER, cm)61W, 54 diagonal
Head coil (DIAMETER, cm)26.8
POWER NEEDED
Line voltage, V (POWER NEEDED)380/400/420/440/480
Kva (POWER NEEDED)40
A/c, btu/hr (POWER NEEDED)45kW peak
Cooling method (X-RAY TUBE)Closed loop, chilled water
Chilled h2o, l/min (POWER NEEDED)5
CRYOGEN USE, L/hr
Liquid helium (CRYOGEN USE, L/hr)<0.1
MAGNET WEIGHT, kg (CRYOGEN USE, L/hr)8900
DIMENSIONS (HXWXD),CM (CRYOGEN USE, L/hr)250x 215 x 228.4
ECHO PLANAR IMAGING (CARDIAC GATING)See Other Specs
AngiographyYes
BORE FEATURES (DIAMETER, cm)Intercom, lighting, nurse-call squeeze bulb; optional music headphones and CCTV
IMAGING
Pulse sequences (IMAGING)SE, IR, GE, FLASH, FISP, PSIF, DESS, TurboFLASH, Turbo Spin Echo, Turbo Inversion Recovery, Turbo Gradient Spin Echo, MP Rage
Repetition time, msec (IMAGING)22-5,000 SE, 3.5minimum GE
Echo time, msec (IMAGING)12minimum SE, 1.7minimum GE
Inversion time, msec (IMAGING)20-20,000
Slice thickness, mm (GANTRY)0.5-20 (2-D), 0.3-62 (3-D)
Fov, cm (IMAGING)2-50
Fov offsets (IMAGING)±25
Scan orientations (IMAGING)Orthogonal, oblique, MAO
Measuring matrix (IMAGING)Up to N x 1024, N = 32 to 1024 in increments of 1
Display matrix (IMAGE DISPLAY)1024x 1024
Pixel intensity (IMAGING)1-4,096
Spatial resolution, mm (IMAGING)0.08,in plane
MAGNETSuperconducting
MARKETING REGION (Interference compensation)Worldwide