Manufacturer Specifications - Mobile MAGNETOM Impact Expert, Siemens

Please note: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.
MODELMAGNETOM Impact Expert
YEAR FIRST INSTALLED 1995
CLINICAL USE Whole body
ELECTROMAGNETIC FIELD STRENGTH, T1
OPTIONSPACSnet Ethernet connection, color monitor, satellite console, Prominence workstation, SatPads for uniform coil load, Advanced coil- positioning kit, additional memory & optical disk drive
CARDIAC GATING
Ecg/peripheral Yes/yes
ECHO PLANAR IMAGING See Other Specs
IMAGING MODES
Single Yes
Multislice Yes
Volume study Yes
GRADIENT SUBSYSTEM
Strength, mt/m 1520 optional
Rise time to 10mT/m, msec <0.6
Cpu Sun-based host integrated into SMI
Memory size, mb 64
ARRAY PROCESSOR Siemens Medical Imager (SMI)
Memory size, mb 288
Storage media/size 5.25" MOD drive; 600MB
Image storage capacity >9,000 (256 x 256)
Respiratory gating Yes
SHIMMING Passive
Diameter, cm
Body coil 60W x 53.5 H, elliptical
Head coil 26.8
AngiographyYes
SURFACE COILS
Spine CP array std
Knee CP trans/rec std
Neck HH volume std
Tmj Bilateral option
Extremity CP Flex pkg, option
Head CP std
Breast Bilateral option
Shoulder CP Flex pkg option
Additional Head/neck MRA, endo- rectal, CP body array and movable spine
BORE DIAMETER or W x H, cm Rectangular, w below 100 cm.60 x 53.5
DICOM 3.0 COMPATIBLE Optional
Reconstruction time
Single slice, sec <0.1
Multislice, sec <0.1
Volume, sec <0.1
BORE FEATURES Intercom, lighting, nurse-call squeeze bulb; optional music headphones and CCTV
5-GAUSS FRINGE FIELD
Radial/axial, m 2.4/4.3
POWER NEEDED
Line voltage, V 380/400/420/440/480
Kva 28
A/c, btu/hr 11.4kW
Cooling method Closed loop, chilled water
Chilled h2o, l/min 5
CRYOGEN USE, L/hr
Liquid helium <0.1
MAGNET WEIGHT, kg 4900
DIMENSIONS 240x 210 x 189
IMAGING
Pulse sequences SE, IR, GE, FLASH, FISP, PSIF, DESS, TurboFLASH, Turbo Spin Echo, Turbo Inversion Recovery, Turbo Gradient Spin Echo, MP Rage
Repetition time, msec 25-2,000 SE, 10minimum GE
Echo time, msec 12minimum SE, 4minimum GE
Inversion time, msec 25-20,000
Slice thickness, mm 0.5-20 (2-D), 0.36-60 (3-D)
Fov, cm 2.5-45
Fov offsets +/-20
Scan orientations Orthogonal, oblique, MAO
Measuring matrix Up to N x 1024, N = 32 to 512 in increments of 1 † 
Display matrix 1024x 1024
Pixel intensity 1-4,096
Spatial resolution, mm 0.2,in plane
MAGNETSuperconducting
You may also be interested in the following Manufacturer Specifications:
Disclaimer

Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.

MedWOW disclaims any obligation, responsibility or liability which concerns or relates to any information on this page.

Mobile MRI, Siemens, Mobile MAGNETOM Symphony