| MODEL: | ECAT EXACT 47 |
| COMPUTER SYSTEM: | |
| Acquisition cpu | Motorola 68030, SPARC processor |
| Storage media/size | Up to 1 GB storage |
| ARRAY PROCESSOR : | 320MFLOPS |
| Image processor | RISC-based workstation |
| Storage media/size | Up to 1 GB storage |
| Display: | 16" high-resolution color |
| Printing options | Laser, color dye sublimation |
| Acquisition possibilities | 2-D, 3-D, static, dynamic, whole body, gated cardiac |
| OPTIONAL ATTACHMENTS | Cardiac gating, isolation trans- former, film hard- copy imager, remote workstations, additional erasable optical disk drives |
| DETECTOR CHARACTERISTICS: | |
| # of rings | 24 |
| Ring diameter, cm | 82 |
| # of crystals | 9216 |
| Crystal composition | BGO |
| Crystal size, mm | 6.75x 6.75 x 20 |
| Axial fov, mm | 162 |
| Crystals/pmt | 16 |
| Patient Table: | |
| Vertical motion,cm | 57-93 |
| Horizontal motion, cm | 200 |
| Head support | 2,fixed and adjustable |
| Arm support | Fixed |
| Max.patient weight capacity, kg | 159(350) |
| INSTALLATION REQUIREMENTS: | |
| Minimum room size | 6x 4 m, scanner room; 3.2 x 4 m, operator room |
| WORK AREA CONDITIONS: | |
| Working temp. c | 22+/-2(72+/-6) |
| Humidity: | 35-70%. noncondensing |
| System cooling btu/hr | 4500 |
| IMAGE RECONSTRUCTION: | |
| Image uniformity | |
| Coincidence window, nsec | 12 |
| Reconstruction time, sec | |
| Maximum count rate, cps @50% dead time | 150,000 (2-D); 350,000 (3-D) |
| Gantry: | |
| H x W x D, cm | 188x 183 x 81 (74x 72 x 31.9) |
| WEIGHT, kg | 1,180(2,601) |
| Patient port diameter, cm | 56.2 |
| PATIENT POSITIONING : | Triple laserand rectilinear scan |
| Transmission source | 3 x Ge-683 Ge-68 rods (with automated extend/retract) |
| Other characteristics | Retractable septa for 3-D acquisition |
| SOFTWARE AVAILABLE | Clinical Applications Programming Package (CAPP), oncology, neurology, cardiology |
| POWER NEEDED: | 230V, 50/60 Hz |
| DETECTOR CAPABILITIES: | |
| System sensitivity, cps/mCi/cc | 180,000 (2-D); 1,400,000 (3-D) |
| Scatter fraction | |
| Resolution, mm | |
| Transaxial fwhm: | |
| 0cm rad, statny | 6 |
| 10cm rad, statny | 6.7 |
| Axial fwhm: | |
| 0cm radius | 5.4 |
| 10cm radius | 6.3 |
| Energy resolution | |
| OTHER ATTRIBUTES | Water cooled for quiet operation; no electronics room required; fiberoptic connection between operator console and gantry; foot extension on patient couch. Meets requirements of CE, CSA, ETL, IEC 601-1, and NEMA performance measurements. |
| MARKETING REGION | Worldwide |