MedWOW / Medical Equipment / Intensive Care Unit / Ventilator, Intensive Care / Viasys Healthcare / Vela / Manufacturer Specifications
Manufacturer Specifications - Vela, Viasys Healthcare
Please note: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.

| MODEL | Vela |
| ALERTS (CONFIGURATION) | |
| Gas supply malfunction (ALERTS) | Yes |
| Power loss (ALERTS) | Yes |
| Vent inoperative (EQUIPMENT ALARMS) | Yes |
| Low battery (EQUIPMENT ALARMS) | 40% and 20% warnings |
| SELF-TEST | Yes |
| Additional (Alarms, high/low) | Test for lamp, leak, switch, alarm, and filter; touchscreen calibration |
| Battery (POWER NEEDED) | Yes |
| Type (number) (ALERTS) | Ni-MH |
| Operating time, hr (Spontaneous/cpap) | 6 |
| Rechargeable (POWER NEEDED) | Internal controller |
| Recharging time, hr (POWER NEEDED) | 8 |
| DIMENSIONS (HXWXD) CM,(IN) (ALERTS) | 30.5 x 33 x 36.8 (12 x 13 x 14.5) |
| WEIGHT, kg (lb) (DISPLAY) | 17.2 (38) |
| Factory/on-site (Service) | Yes/yes |
| TRAINING (SYSTEM CONFIGURATION) | Class at factory |
| 3rd-party service (Service) | In some areas |
| Parts availability (ALERTS) | Yes |
| Delivery time, aro (SHOCK-WAVE GENERATOR) | <60 days |
| Year first sold (PURCHASE INFORMATION) | 2002 |
| Usa/worldwide | |
| Fiscal year (SHOCK-WAVE GENERATOR) | January to December |
| REGULATORS | |
| Tidal volume, ml (CONTROLS) | 50-2,000 |
| Insp flow, l/min (REGULATORS) | 10-140/180 spont max |
| Insp press, cm h2o (REGULATORS) | 0-100 |
| Respiratory rate, breaths/min (REGULATORS) | 29252 |
| Insp time, sec (CONTROLS) | 0.3-10 |
| Exp time, sec (REGULATORS) | Depends on rate |
| Insp hold/plateau (REGULATORS) | Off, 0.1-2 sec |
| Expiratory hold (REGULATORS) | Manual |
| Fio2, % (CONTROLS) | 21-100 |
| Manual breath (REGULATORS) | Yes |
| Peep/cpap, cm h2o (REGULATORS) | 0-35 |
| Pressure support, cm h2o (REGULATORS) | Off, 1-60 |
| Nebulizer (REGULATORS) | Off, 1 to 60 min |
| Trigger system (REGULATORS) | Flow w/press backup |
| Bias/base flow range, l/min (REGULATORS) | 44105 |
| Sigh (REGULATORS) | Yes |
| 100% o2 (REGULATORS) | Yes |
| MONITORED PARAMETERS | |
| Pip (MONITORED PARAMETERS) | Yes |
| Map (MONITORED PARAMETERS) | Yes |
| Peep (MONITORED PARAMETERS) | Yes |
| Tidal (Volume) | Yes |
| Minute (Volume) | Yes (VTI, VTE) |
| Spontaneous minute (Volume) | Yes |
| Fio2 (PATIENT ALARMS) | Yes |
| Respiratory rate (Volume) | Yes |
| Inspiration time (MONITORED PARAMETERS) | Yes |
| Expiration time (MONITORED PARAMETERS) | Yes |
| I:e (CONTROLS) | Yes |
| Additional (Alarms, high/low) | High and low PIP, high rate, safety valve open, spont & mandatory minute and tidal volumes, Ti (sec), Te (sec) |
| POWER NEEDED, VAC (ENERGY) | 90-240, 47-65 Hz |
| Current, amps (REGULATORS) | 2 |
| POWER CONSUMPTION, W (OTHER AVAILABLE ROTORS) | 36-220 |
| DATA DISPLAYED (REGULATORS) | User configurable |
| PATIENT ALERTS O2 (REGULATORS) | Yes |
| Low minute volume (REGULATORS) | Yes |
| Low inspir pressure (REGULATORS) | Yes |
| Over pressure (MONITORED PARAMETERS) | Yes |
| Loss of peep (EQUIPMENT ALARMS) | Low pressure |
| Apnea (CONFIGURATION) | Yes |
| High continuous pressure/occlusion (EQUIPMENT ALARMS) | Circuit fault |
| High respir rate (REGULATORS) | Yes |
| High minute vol (REGULATORS) | Yes |
| High peep (PATIENT ALARMS) | Yes |
| PNEUMATIC POWER (Volume) | |
| Compressor | Batt backed turbine |
| Compressed gases (EQUIPMENT ALARMS) | O2 |
| Pressure ranges (EQUIPMENT ALARMS) | 40-80 psig |
| INTERFACING | |
| Output ports (INTERFACING) | RS232 |
| Remote alrm/display (INTERFACING) | Nurse call, others |
| ANALOG OUTPUT (CONFIGURATION) | Direct video |
| DISPLAY | LCD built-in, SVGA |
| Hard copy (Report generation) | HP 940 driver output |
| Network (Report generation) | BGM option |
| DISPLAYS (STAINING POSTS) | Graphics SVGA color LCD |
| MODES OF OPERATION | |
| Volume breaths (Assist/control) | Yes |
| Pressure breaths (Assist/control) | Yes |
| Volume breaths (Assist/control) | Yes |
| Pressure breaths (Assist/control) | Yes |
| Pressure support (Simv) | Yes |
| Pressure support (Simv) | Yes |
| Apnea-backup vent (Spontaneous/cpap) | Yes |
| Additional (Alarms, high/low) | Volume assist control with Vsync |
| OTHER SPECIFICATIONS (AIR SUPPLY) | Six-hour battery; low pressure (flowmeter); O2 inlet connection; synchronous nebulizer drive; touchscreen controls; transport cart option with cylinder holders. |
| FDA CLEARANCE (Interference compensation) | Yes |
| CE MARK (MDD) (Interference compensation) | Yes |
| WHERE MARKETED | Worldwide |
You may also be interested in the following Manufacturer Specifications:
Disclaimer
Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.
Ventilator, Intensive Care, Viasys Healthcare, Vela
![]() | View More Items | ![]() |