Manufacturer Specifications - 760, Mallinckrodt Puritan Bennett

Please note: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.
MODEL760
ALERTS (CONFIGURATION)
Gas supply malfunction (ALERTS)Yes
Power loss (ALERTS)Yes
Vent inoperative (EQUIPMENT ALARMS)Yes
Low battery (EQUIPMENT ALARMS)Yes
SELF-TESTElectronics and pneumatics
Additional (Alarms, high/low)Disconnect, occlusion
Battery (POWER NEEDED)Yes
DIMENSIONS (HXWXD) CM,(IN) (ALERTS)37.8 x 51.5 x 37 (14.9 x 20.3 x 14.6) without cart 1
WEIGHT, kg (lb) (DISPLAY)30 (66.2) ventilator, 18 (39.7) cart 125.5 x 51.5 x 37 cm (49.4 x 20.3 x 14.6 in) withcart.
Factory/on-site (Service)Yes/yes
TRAINING (SYSTEM CONFIGURATION)Yes
3rd-party service (Service)Yes
Parts availability (ALERTS)To customers, others
Delivery time, aro (SHOCK-WAVE GENERATOR)45 days
Year first sold (PURCHASE INFORMATION)1998
Usa/worldwide
Fiscal year (SHOCK-WAVE GENERATOR)July to June
REGULATORS
Tidal volume, ml (CONTROLS)40-2,000
Insp flow, l/min (REGULATORS)3
Insp press, cm h2o (REGULATORS)29342
Respiratory rate, breaths/min (REGULATORS)25569
Insp time, sec (CONTROLS)0.2-8
Exp time, sec (REGULATORS)>0.2
I:e ratio (REGULATORS)1:99 to 4:1
Insp hold/plateau (REGULATORS)0-2 sec
Fio2, % (CONTROLS)21-100
Manual breath (REGULATORS)Yes
Peep/cpap, cm h2o (REGULATORS)0-35
Nebulizer (REGULATORS)Ultrasonic
Trigger system (REGULATORS)Flow
Pressure slope/ramp adjustment (REGULATORS)Yes
POWER NEEDED, VAC (ENERGY)100-240
Current, amps (REGULATORS)2.9 (115 V)
POWER CONSUMPTION, W (OTHER AVAILABLE ROTORS)290
MONITORED PARAMETERS
Pip (MONITORED PARAMETERS)Yes
Map (MONITORED PARAMETERS)Yes
Peep (MONITORED PARAMETERS)Yes
Tidal (Volume)Yes
Minute (Volume)Yes
Spontaneous minute (Volume)Yes
Fio2 (PATIENT ALARMS)Yes
Inspiration time (MONITORED PARAMETERS)Yes
Expiration time (MONITORED PARAMETERS)Yes
I:e (CONTROLS)Yes
Additional (Alarms, high/low)E-sense
DISPLAYS (STAINING POSTS)LCD
PNEUMATIC POWER (Volume)
Compressed gases (EQUIPMENT ALARMS)O2
Pressure ranges (EQUIPMENT ALARMS)40-90 psi
INTERFACING
Output ports (INTERFACING)RS232 (2)
Remote alrm/display (INTERFACING)Yes
DISPLAYOptional
Hard copy (Report generation)Yes
Archival disk (Report generation)Yes
Network (Report generation)Yes
DATA DISPLAYED (REGULATORS)Numbers, optional graphics
MODES OF OPERATION
Volume breaths (Assist/control)Yes
Pressure breaths (Assist/control)Yes
Volume breaths (Assist/control)Yes
Pressure breaths (Assist/control)Yes
Pressure support (Simv)Yes
Apnea-backup vent (Spontaneous/cpap)Yes
Additional (Alarms, high/low)Occlusion cycling, disconnect
OTHER SPECIFICATIONS (AIR SUPPLY)Exhalation sensitivity; rise time percentage; integral O2 sensor; 2-5 hr (optional 7 hr) battery backup. Meets requirements of CISPR 11; CSA; IEC 60601-1, 60601-2, and 60601-2-12; and UL.
FDA CLEARANCE (Interference compensation)Yes
CE MARK (MDD) (Interference compensation)Yes
WHERE MARKETEDWorldwide
You may also be interested in the following Manufacturer Specifications:
Disclaimer

Information on this page may be partial or incomplete and may contain mistakes or inaccurate information.

MedWOW disclaims any obligation, responsibility or liability which concerns or relates to any information on this page.

Ventilator, Intensive Care, Mallinckrodt Puritan Bennett, 760